S3 · 01/21
2026 ·
Auris Health (J&J MedTech)
| Company | Auris Health (J&J MedTech) |
| Category | medical robotics |
| Type | recall |
| Date | 01/21 2026 |
| Time of day | not documented |
| Location | no physical site (recall / fleet-wide / aggregate record) |
| Severity | S3 (recall, enforcement, or fleet-wide fault) |
| Scale | MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005- |
| Confidence | high |
| Verification | Yes - adversarially checked against the cited source |
| Source | openFDA enforcement |
FDA enforcement record (Class II, recall Z-1383-2026): Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may