S3 · 07/2026 · Intuitive Surgical
| Company | Intuitive Surgical |
| Category | medical robotics |
| Type | injury |
| Date | 07/2026 |
| Time of day | not documented |
| Location | no physical site (recall / fleet-wide / aggregate record) |
| Severity | S3 (recall, enforcement, or fleet-wide fault) |
| Scale | MDR 2955842-2026-30540 |
| Confidence | medium |
| Verification | Yes - adversarially checked against the cited source |
| Source | openFDA MAUDE |
INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE UNIVERSAL SEAL FOR FAILURE ANALYSIS EVALUATION. THE SEAL WAS ANALYZED AND FOUND TO HAVE TEARS AT THE RUBBER SIDE OF THE SEAL, MEASURING APPROXIMATE 0.2590" IN LENGTH.
| Sev | Date | Company | Event |
|---|---|---|---|
| S2 | 2015-06-30 | Hocoma | Proximal tibia fracture in a patient with incomplete spinal cord injury associated with robotic treadmill training |