S3 · 07/2026 · Intuitive Surgical
| Company | Intuitive Surgical |
| Category | medical robotics |
| Type | injury |
| Date | 07/2026 |
| Time of day | not documented |
| Location | no physical site (recall / fleet-wide / aggregate record) |
| Severity | S3 (recall, enforcement, or fleet-wide fault) |
| Scale | MDR 2955842-2026-30253 |
| Confidence | medium |
| Verification | Yes - adversarially checked against the cited source |
| Source | openFDA MAUDE |
INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE VESSEL SEALER CURVED INSTRUMENT TO PERFORM FAILURE ANALYSIS. THE INSTRUMENT WAS ANALYZED AND FOUND TO HAVE THE LOWER GRIP BROKEN. THREE PIECES MEASURING APPROXIMATELY 8.28MM X 11.77MM, 3.70MM X 7.02MM AND 1.91MM X 8.02MM IN SIZE WERE BR