S3 · 09/2022 · Intuitive Surgical
| Company | Intuitive Surgical |
| Category | medical robotics |
| Type | injury |
| Date | 09/2022 |
| Time of day | not documented |
| Location | no physical site (recall / fleet-wide / aggregate record) |
| Severity | S3 (recall, enforcement, or fleet-wide fault) |
| Scale | MDR 2955842-2022-13942 |
| Confidence | medium |
| Verification | Yes - adversarially checked against the cited source |
| Source | openFDA MAUDE |
INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE SYNCHROSEAL INSTRUMENT INVOLVED WITH THIS COMPLAINT AND COMPLETED THE DEVICE EVALUATION. FAILURE ANALYSIS CONFIRMED THE REPORTED COMPLAINT. THE INSTRUMENT PIVOT PIN WASHER WAS NO LONGER ATTACHED TO THE JAW COVER AT THE WRIST ASSEMBLY. T