S3 · 04/2024 · Intuitive Surgical
| Company | Intuitive Surgical |
| Category | medical robotics |
| Type | injury |
| Date | 04/2024 |
| Time of day | not documented |
| Location | no physical site (recall / fleet-wide / aggregate record) |
| Severity | S3 (recall, enforcement, or fleet-wide fault) |
| Scale | MDR 2955842-2024-13071 |
| Confidence | medium |
| Verification | Yes - adversarially checked against the cited source |
| Source | openFDA MAUDE |
THE VIDEO PROCESSOR (VP) UNIT WAS RETURNED AND ANALYZED, AND THE REPORTED FAILURE WAS REPLICATED. THIS UNIT WAS INSTALLED INTO A TEST SYSTEM. DURING THE TEST, THE ISSUE WAS REPLICATED. DURING TROUBLESHOOTING, THE FANS WERE NOT WORKING. THE CAUSE WAS THE VIDEO PROCESSOR POWER DIST