S3 · 05/2023 · Intuitive Surgical
| Company | Intuitive Surgical |
| Category | medical robotics |
| Type | injury |
| Date | 05/2023 |
| Time of day | not documented |
| Location | no physical site (recall / fleet-wide / aggregate record) |
| Severity | S3 (recall, enforcement, or fleet-wide fault) |
| Scale | MDR 2955842-2023-13460 |
| Confidence | medium |
| Verification | Yes - adversarially checked against the cited source |
| Source | openFDA MAUDE |
INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE HARMONIC INSERT INVOLVED WITH THIS COMPLAINT AND COMPLETED THE DEVICE EVALUATION. THE REPORTED COMPLAINT WAS CONFIRMED DURING FAILURE ANALYSIS. THE HARMONIC INSERT WAS FOUND TO HAVE A BROKEN BLADE. THE BLADE BROKE AT ROUGHLY .232¿ FROM