S3 · 04/2025 · Intuitive Surgical
| Company | Intuitive Surgical |
| Category | medical robotics |
| Type | injury |
| Date | 04/2025 |
| Time of day | not documented |
| Location | no physical site (recall / fleet-wide / aggregate record) |
| Severity | S3 (recall, enforcement, or fleet-wide fault) |
| Scale | MDR 2955842-2025-13347 |
| Confidence | medium |
| Verification | Yes - adversarially checked against the cited source |
| Source | openFDA MAUDE |
INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE SYNCHROSEAL INSTRUMENT WAS ANALYZED. THE INSTRUMENT WAS FOUND TO HAVE SPRING PAD DAMAGE ON THE ELECTRODE CAUSING IT TO BECOME STUCK TO THE BLUE SPRING PAD ON THE OPPOSITE SIDE OF THE JA