S3 · 06/2023 · Intuitive Surgical
| Company | Intuitive Surgical |
| Category | medical robotics |
| Type | injury |
| Date | 06/2023 |
| Time of day | not documented |
| Location | no physical site (recall / fleet-wide / aggregate record) |
| Severity | S3 (recall, enforcement, or fleet-wide fault) |
| Scale | MDR 2955842-2023-15442 |
| Confidence | medium |
| Verification | Yes - adversarially checked against the cited source |
| Source | openFDA MAUDE |
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED COMPLAINT. THE REPORTED COMPLAINT WAS REPRODUCED DURING FIELD EVALUATION. THE FSE REVIEWED THE LOG AND FOUND ERRORS 23002 AND 23008 POINTING TO THE RIGHT MTM, AXIS 8. THE FSE TR
| Sev | Date | Company | Event |
|---|---|---|---|
| S2 | 2015-06-30 | Hocoma | Proximal tibia fracture in a patient with incomplete spinal cord injury associated with robotic treadmill training |