S3 · 01/2025 · Intuitive Surgical
| Company | Intuitive Surgical |
| Category | medical robotics |
| Type | injury |
| Date | 01/2025 |
| Time of day | not documented |
| Location | no physical site (recall / fleet-wide / aggregate record) |
| Severity | S3 (recall, enforcement, or fleet-wide fault) |
| Scale | MDR 2955842-2024-24084 |
| Confidence | medium |
| Verification | Yes - adversarially checked against the cited source |
| Source | openFDA MAUDE |
INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE HARMONIC ACE INSTRUMENT WAS ANALYZED AND FOUND TO HAVE THE CURVED BLADE BROKEN AT THE TIP/MIDPOINT/BASE. A PIECE APPROXIMATELY 0.580" X 0.087" WAS FOUND TO BE BROKEN OFF. THE BROKEN PIE
| Sev | Date | Company | Event |
|---|---|---|---|
| S2 | 2015-06-30 | Hocoma | Proximal tibia fracture in a patient with incomplete spinal cord injury associated with robotic treadmill training |